17 July 2000

Hajime Kawakami

Life time performance of FM-intensifier in analog mode

 
 

1. Introduction

The previous report (Swift-UVOT/MSSL.TC/0002) showed capability of analog mode for high time resolution photometry, when a target star was brighter than 17.4mag. Swift UVOT may have several bright stars in the field of view during monitoring time variation of a gamma ray burster. It is dangerous for the intensifier to observe a bright star for a long time in photon counting mode (high gain in MCPs). The analog mode observation will offer the longer life time, since the electric gain of MCPs is less than 1/6 of the photon counting mode.

Quantifying very bright star as a stand alone system is one of the safety requirements for Swift UVOT. Analog mode is one of the candidate to fulfil this requirement. It is, however, essential to know, how long does the detector withstand against the intense star illumination during the brightness assessment.

Our intensified CCD detector demonstrated far longer life time (>100 times) than typical position sensitive detectors in terms of accumulated anode charge (XMM-OM/MSSL/TC/ 0059). This difference may be due to the lower gain of MCPs with our detector, i.e. ~5x10^5 with ours while ~10^7 with the position sensitive detectors. If the life time depends on the gain strongly, the reduction of the gain by the factor of 6 may extend the life time far longer than x6.

In this report, the image intensifier was operated in the analog mode and was exposed to intense pinhole illuminations for 100 hours. Gain depletion of MCPs and sensitivity loss in F-F images were investigated against accumulated charge.
 
 

2. Electric gain

The best result in analog mode was obtained with 90% of nominal photon counting MCP voltage in the test with DEP_#5 intensifier. The experiments in this report was carried out with DEP_#8 intensifier, whose nominal MCPs voltage is 2250V. The 2020V, 90% of the nominal, was applied to the MCPs during the photon dose. The illumination pattern was a 11x11 pinhole array and the brightness has gradient along the column in the dynamic range of ~2E+4 (Fig. 1). The light source is made of 64 green LEDs covered with a diffuser and a 5300-5700A interference filter. The brightness of the LEDs is controlled by a constant current source in the dynamic range of ~2E+4 (see detail in XMM-OM/MSSL.TC/0057).
 
 

 The brightness of the pinhole array was calibrated with3 photon counting images in the 3 LED current levels, L=1, L=3 and L=10 in order to overcome small dynamic range of the detector. The lower LED current (L=1) was used for determining brightness ratio among bright pinhole columns, while the medium LED current (L=3) was for faint pinhole columns (Table 1). The highest current (L=10) was for the calibration of LED brightness during the photon dose, in which the faintest pinhole column (col=1) still gave useful data. Photon losses due to coincidence were corrected and the true input rates were tabulated in Table 2. Finally, the absolute brightness of the pinhole columns at the LED current level of L=10 was tabulated in Table 3.

 

Table 1. Raw counts /(hour x spot) 21 June 2000 DEP_#8

-------------------------------------
LED =                   1                      3                        10
-------------------------------------
col=11         264920.0               N/A                   N/A
col=10         254720.0               N/A                   N/A
col=  9           32620.0                N/A                   N/A
col=  8           24460.0                N/A                   N/A
col=  7             2381.2                N/A                   N/A
col=  6             2131.9         23104.0                    N/A
col=  5               158.2            1680.4                    N/A
col= 4                 84.1            1042.4                    N/A
col=  3                 22.5              186.2                    N/A
col=  2                ( 7.0)             169.0             (253160.0)
col=  1                 14.2              164.6               264110.0
-------------------------------------
Table 2. True counts /(sec x spot) 21 June 2000 DEP_#8
-------------------------------------
LED =                   1                      3                        10
-------------------------------------
col=11        94.58122
col=10        90.11536
col=9             9.74329
col=8             7.26678
col=7               .69737
col=6               .62426             6.85783
col=5               .04627               .49191
col=4               .02459               .30502
col=3               .00658               .05446
col=2             (.00205)              .04942           (89.43997)
col=1               .00415               .04814             94.22338
-------------------------------------
Table 3. Pinhole brightness at LED current level = 10
---------------------------------------------------------------
column           1           2           3           4           5           6           7           8           9           10           11
---------------------------------------------------------------
Brtness       94.22    96.73  106.59   597.0   962.8   13.4k   15.0k   156k    209k     1938k    2034k
(c/s)
B0 star         16.5      16.5      16.4      14.5     14.0     11.1     11.0       8.4       8.1          5.7         5.6
(mag)
---------------------------------------------------------------
The ratio of gains between V_mcp=2250V(photon counting) and 2020V(analog) were determined from both of the brightness of event splash at phosphor screen and anode current. The pulse height distributions of the event splash with the 2 different voltages to MCPs were shown in Fig. 2. Flat filed in the count rate of 15,000 c/(sec full area) were used for the illumination. The brief ratio was determined from the peak positions, > 5.5 times. It was difficult to determine the ratio accurately, because the pulse height distribution with V_mcp=2020V was squashed to the lower energy end.



The brighter F-F illumination was used for the measurement of the anode current to produce sufficient current with V_mcp=2020V. The detected count rate for the F-F was 86,100 c/(sec full area) in photon counting mode. After the correction of the coincidence loss, the true incoming rate was turned out to be 94,000 c/(sec full area). The procedure of the coincidence correction followed XMM-OM/MSSL.TC/0050. Coincidence area of event splashes was assumed to be 12 (CCD_pixels)^2 from other 2 intensifiers, though there was no data on DEP_#8 itself. The full detector area in the photon counting imaging is (3.37 x 256 CCD_pixels)^2, while photocathode area is circle with the diameter of 25mm. Since the anode current was produced from all photocathode area, incoming rate of electrons involved in anode current was 94,000 c/s * (D=25mm) / (3.37 x256 (CCD_pixels)^2 = 94,000 c/s * 1.2467 = 117,000 c/s. A 99.91k Ohm resister was inserted at the anode cable, whose voltage was at 8000V, and the small voltage drop across the resister was measured with a precision multimeter, FLUKE 87 IV, in the minimum readout of 1uV. The resistance value was also calibrated by the FLUKE 87 IV. The small voltage drops were 1012uV and 151uV with V_mcp=2250V and 2020V. Hence, the currents were 10.23nA and 1.53nA. Since the input impedance of Fluke 87IV was 10M Ohm, anode currents were corrected by the factor of 1.01. Finally, the electric gain was calculated to be 5.4x10E+5 with V_mcp=2250V and 8.1x10E+4 with V_mcp=2020V. The ratio of the gains with the 2 voltages was x6.7 times.
The electric gain in the high input rate was measured using pinhole illuminations in the LED current levels of L=1-10 for V_mcp=2250V and L=3-10 for V_mcp=2020V. Columns=1-9 of the pinhole array was blocked for this measurement, so that the brightest 2 columns=10-11 with nearly same brightness were used for the illumination. Since voltage display of the FLUKE 87 IV was not stable in the last 2 digits (10uV, 1uV), the display was read 10 times and was averaged for the lowest 2 illuminations (i.e. LED current levels L=3 and L=4 for V_mcp=2020V and L=1 and L=2 for V_mcp=2250V). The results for the both of V_mcp=2250V and V_mcp=2020V were tabulated in Table 4 and were plotted in Figs. 3 and 4. The electric gain of the intensifier was 5.7E+5 in the low input rate with V_mcp = 2250V and 8.1E+4 with V_mcp = 2020V. The gain depletion is 1/9.7 in the input rate of 2E+6 c/s with V_mcp = 2250V, while 1/8 with V_mcp = 2020V, compared with those at the input rate of 100 c/s. The electric gains of pinhole illumination in the low input rate was higher than that of F-F illumination. This was due to global gain variation of MCPs ( i.e. the local gain at pinhole positions were higher than the average).
Electric gain of MCPs at pinhole positions should have changed during the photon dose. The anode current was measured after completing the 100 hours photon dose by illuminating exactly same positions. This gauges the level of the change before and after the photon dose. Again, columns=1-9 of the pinhole array was blocked for the measurement, so that the brightest 2x11 pinholes from columns=10-11 were used for the illumination. The gains at the brightest pinhole positions at different input rate were tabulated in Table 5 and were plotted in Figs. 3 and 4 overlaying on the original gains. In spite of the large gain loss in the low input rate, the gain in the high input rate does not change before and after the 100 hours dose. This is particularly true for the illumination above 1E+5 c/(sec x spot) with V_mcp = 2250V and 1E+6 c/(sec x spot) with V_mcp = 2020V. From these results, we can assume anode currents at columns=10 and 11 were constant throughout the photon dose, hence we can estimate total accumulated charge precisely. There is no measurement on the change of gain at other pinhole positions, i.e. columns=1-9. Since the total accumulated charges are smaller, the anode currents hopefully did not change much before and after the photon dose.

Because of the large gain loss in the low input rate while no gain loss in the high input rate after the 100 hours photon dose, the gradient of the gain curve against input rate becomes flat. This suggests a very hard scrubbing may lighten the effect of pore paralysis, hence may extend dynamic range of MCPs.
 

Table 4. Electric gain of XMM-OM tube in high count rate

--------------------------------------------------------
              LED Intensity          Anode current (pA)               Electric Gain
              (c/s pinhole)              from 22 pinholes
                                                  2020V         2250V                2020V        2250V
--------------------------------------------------------
F-F        94000                         1530           10230                8.1 E+4         5.4 E+5
L=1          92.35                          (6.7)              184                (2. E+4)        5.7 E+5
L=2             352                           132               585                10. E+4          4.7 E+5
L=3           1014                           295             1370                8.3 E+4          3.8 E+5
L=4           3426                           800             2880                6.6 E+4          2.4 E+5
L=5         16500                         2330             9260               4.0 E+4          1.6 E+5
L=6         51000                         5250           20200               2.9 E+4          1.1 E+5
L=7       139000                       11100          43300               2.3 E+4         0.88E+5
L=8       410000                       29700        114000               2.1 E+4         0.79E+5
L=9       984000                       57800        225000               1.7 E+4         0.65E+5
L=10   1986000                    102000        412000               1.5 E+4         0.59E+5
--------------------------------------------------------
 

Table 5. Electric gain after 100 hours dose

--------------------------------------------------------
              LED Intensity          Anode current (pA)               Electric Gain
              (c/s pinhole)              from 22 pinholes
                                                  2020V         2250V                2020V        2250V
--------------------------------------------------------
L=1          92.35                          ---                 64                   ---              2.0 E+5
L=2             352                          ---               219                   ---              1.8 E+5
L=3           1014                          103               664                  3. E+4           1.9 E+5
L=4           3426                          239             1810                  2. E+4           1.5 E+5
L=5         16500                        1041             6550               1.8 E+4           1.1 E+5
L=6         51000                        2790           16900               1.6 E+4          0.94E+5
L=7       139000                        7700          38400                1.6 E+4         0.78E+5
L=8       410000                      24000        108000               1.6 E+4          0.75E+5
L=9       984000                      50200        225000               1.4 E+4          0.65E+5
L=10   1986000                      94100        419000               1.3 E+4         0.60E+5
-------------------------------------------------------------------
Ref-2 Files used in this section
/swift/ZPHD010.dat
ZPIN011.dat,ZPIN012.dat,ZPIN013.dat,ZPIN014.dat

 
 

3. Gain loss of MCPs

A reference pulse height distributions for individual pinhole columns, col=4-11 (600-2E+6 c/s), were measured with V_mcp=2250V (photon counting mode) before starting the photon dose. The photon doses were carried out 3 times for 15 hours, 15 hours and 70 hours with V_mcp=2020V. The pulse height distributions were measured after the each photon dose. Fig. 5 shows the reference pulse height distribution and the one after the 100 hours dose by the 2E+6 c/s pinholes.




The gain reduced to 1/2.5 of the beginning. The gain loss was quantified from peak positions of the pulse height distributions. The gains after each dose were tabulated in Table 6, and were plotted against accumulated charge in Fig 6. Fig. 7 is the extract from XMM-OM/MSSL.TC/ 0059, in which the intensifier was operated in photon counting mode during the photon dose. The plots in analog mode coincides with those in photon counting mode. This implies that the gain loss can be described by the single parameter, accumulated charge, in any condition (i.e. different gain, input count rate, exposure time etc.).




Table 6. Gain depletion of MCPs

----------------------------------------------------------
  Dose                                              Pinhole intensity (counts/sec)
     time            2030k     1940k        210k       160k          15k          13k          960          600
----------------------------------------------------------
  15.0 hr            .623         .609         .735         .816         .890         .884         .907         .941
  30.0 hr            .570         .573         .772         .807         .884         .862         .922         .916
100.0 hr            .363         .346         .589         .570         .793         .811         .924         .955
----------------------------------------------------------

 

Ref-3 Files used in this section

/swift/ZPHD016.dat,ZPHD028.dat,ZPHD047.dat,ZPHD064.dat

 

4. Sensitivity loss in photon counting image

A reference F-F image with the blue (460nm) LED illumination was integrated for 15 hours in photon counting mode before starting the photon dose. F-F image integrations were followed after each intense illumination to see the sensitivity losses in the different level of dose. The integrations started after the sufficient period since the end of the intense illumination in order to avoid fluorescence, i.e. 80 hours in the 1st day, 38 hours the 2nd day and 27 hours the last day.

Fig. 8 shows 4 F-F images, one was taken prior to the photon dose for reference and the other 3 were after 15, 30 and 100 hours dose. The F-F after 100 hours dose clearly showed an array of black spots corresponding to the pinhole positions.





A F-F image in each day of photon dose was divided by the reference F-F to remove detector artefacts and illumination non-uniformity. Then, the 11x11 array of black spots were averaged along the columns to improve S/N. Central positions of the black spots coincided with the pinhole positions in the accuracy of 8um along H_intensifier direction (bias direction of the 1st MCP), while systematically shifted by 15um along V_intensifier direction. The day by day growth of the black spots is shown in Fig. 9. The image contains all factors, i.e. fluorescence, gain depletion and photocathode sensitivity loss. Fig. 10 shows profiles of the averaged black spots in the last day. Y-width of the slice is 3 twixel (= 58um). The depth of black spots reached 25% at the brightest pinhole after 100 hours dose. The black spots were noticeable for the illumination intensities of > 13kc/s after 100 hours dose, but not for the lower intensity illuminations. Since the F-F integrations were started after the sufficiently long period since the end of the photon dose, the peak depths were only little affected by fluorescence (less than 1.4%).


 
 




The 6-10 dark frames were acquired in photon counting mode before and after the F-F integrations in order to correct the effect of fluorescence for further precision. Standard fluorescence profiles are shown in the lower panel of Fig. 9. The size of the fluorescence is far larger than both of the black spots and projected pinhole, i.e. from 220um to 270um. The fluorescence spots have offset from the pinhole positions by 200-240um along H_intensifier direction (bias direction of the 1st MCP). The fluorescence spot got doughnut shape (i.e. black spot in central part) with the increase of the dose level. The details of these characteristics were identical to those described in the former report, XMM-OM/MSSL/TC/0057.

The sensitivity loss at the peak position was quantified from the average of 3x3 twixels square centred on the black spots. The normalization level was determined from 37x37 twixels (=717um) square excluding central D=21 twixels circular area. Then, the effect of fluorescence (1.4 in maximum) was subtracted. The results were tabulated in Table 7 and were plotted against accumulated charge in Fig. 11. The sensitivity did not decrease till 1E-5 couloms/spot. It started to decrease steeply from 1E-4 couloms/spot.
 
 




The sensitivity loss seen in F-F image is the combination of gain depletion and photocathode sensitivity loss. The photocathode sensitivity losses were calculated by removing the effect of gain depletion (threshold=15ADU). The results were tabulated in Table 8 and were shown in Fig. 12. The sensitivity loss of photocathode is not noticeable up to 3E-5 couloms/spot.




The photocathode sensitivity loss for the same intensifier but dosed in photon counting mode was extracted from XMM-OM/MSSL/TC/0057 (Fig. 13). The sensitivity loss dosed in analog mode shows slightly faster decrease above 2E-4 couloms/spot. This result did not imply that ion barrier characteristics of the 1st MCP improved with the lower voltage.




A Gaussian profile was fitted to the black spots to quantify the spatial extent. The results are shown in Fig. 14. There are clear correlation with accumulated charge. The width increased from 80um(FWHM) to 110um after acquiring 1E-5 to 2E-3 couloms/spot.





Table 7. Sensitivity change in blue F-F image at peak position

----------------------------------------------------------------------
  Dose                                              Pinhole intensity (counts/sec)
     time            2030k     1940k        210k       160k          15k          13k          960          600          110            97
----------------------------------------------------------------------
  15.0 hr            .933         .942        1.000        .993       1.002        1.010      1.006       1.001          .999        .984
  30.0 hr            .864         .892          .971        .958         .979          .995         .994       1.002        1.000      1.001
100.0 hr            .743         .755          .907        .910         .956          .977         .997         .999        1.001        .997
-----------------------------------------------------------------------------------------------------

 

Table 8. Photocathode sensitivity change at peak position

----------------------------------------------------------------------
  Dose                                              Pinhole intensity (counts/sec)
     time            2030k     1940k        210k       160k          15k          13k          960          600          110            97
----------------------------------------------------------------------
  15.0 hr            .951         .965        1.032      1.009       1.007       1.016       1.014       1.006          .999         .984
  30.0 hr            .886         .917          .994        .975         .985        1.001       1.001       1.009        1.000      1.001
100.0 hr            .801         .823          .955        .950         .966          .983       1.004       1.003        1.001        .997
----------------------------------------------------------------------
Ref-4 Files used in this section
/swift/ZDEP015.dat,ZDEP030.dat,ZDEP043.dat,ZDEP060.dat
ZPHD016.dat,ZPHD028.dat,ZPHD047.dat,ZPHD064.dat
ZDRK019.dat,ZDRK020.dat,ZDRK021.dat,ZDRK029.dat,ZDRK031.dat
ZDRK032.dat
ZDRK033.dat,ZDRK034.dat,ZDRK035.dat,ZPHD036.dat,ZPHD037.dat
ZDRK038.dat,ZDRK039.dat,ZDRK042.dat,ZPHD044.dat,ZPHD045.dat
ZDRK048.dat,ZDRK049.dat,ZDRK050.dat,ZPHD051.dat,ZPHD052.dat
ZDRK053.dat,ZDRK054.dat,ZDRK055.dat,ZPHD059.dat,ZPHD061.dat

 

5. Summary

i) Gain loss dosed in analog mode was same as that in photon counting mode in terms of accumulated anode charge.

ii) Photocathode sensitivity loss dosed in analog mode was slightly faster (by ~30%) than that in photon counting mode in terms of accumulated anode charge.

iii) Gain loss measured in the low input rate was 1/2 - 1/3 after 100 hours photon dose, while that in the high input rate almost nothing. In the consequence, pore paralysis curve was flatten after the dose. This suggests a possibility of extending dynamic range of MCPs by a hard scrubbing.
 
 
 
 

Appendix. Experiment procedure for DEP_#8 intensifier in analog mode

20 June - 6 July 2000
-------------------------------------------------------------------
   File Name                   Pinhole                   PHD                   Dark                   F-F                   Time(start)
-------------------------------------------------------------------
Before Damage for reference
2000/06/20
PHD010                                                           300FRs                                                                     17H 51M 05S
Pin011  L=3                  54000S                                                                                                        19H 01M 34S
2000/06/21
Pin012  L=1                    3600S                                                                                                         10H 17M 16S
Pin013  L=10                  1800S                                                                                                         15H 11M 25S
Pin014  L=10                  1800S                                                                                                         16H 21M 36S
DEP015                                                                                                                 54000S                 18H 13M 44S
2000/06/22
PHD016                                                        70000FRs                                                                     11H 27M 18S
Ana017                                                                                                                 10000FRs             12H 37M 00S
Ana018                                                                                                                 10000FRs             16H 50M 13S
\/\/\/\/\/\/\\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/
                            15 hour                   Day-1                   18:24 - 09:24                                                  2000/06/22
\/\/\/\/\/\/\//\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\\/\/\/\/\/\/\\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\//\/\/\/\/\/\/\/\/
2000/06/23
Drk019                                                                                        7200S                                              10H 36M 54S
Drk020                                                                                        7200S                                              12H 37M 18S
Drk021                                                                                        7200S                                              14H 37M 42S
DEP022                                                                                                    Th=15 54000S                 17H 48M 35S
2000/06/24
Drk023                                                                                        7200S                                              08H 48M 59S
Drk024                                                                                        7200S                                              10H 49M 22S
Drk025                                                                                        7200S                                              12H 49M 45S
Drk026                                                                                        7200S                                              14H 50M 08S
2000/06/26
Ana027                                                                                                                 20000FRs             12H 28M 17S
PHD028                                                        70000FRs                                                                     14H 34M 34S
Drk029                                                                                        7200S                                              15H 37M 46S
DEP030                                                                                                   Th=15 54000S                  17H 38M 10S
2000/06/27
Drk031                                                                                        7200S                                              08H 38M 34S
Drk032                                                                                        7200S                                              10H 38M 58S
\/\/\/\/\/\/\\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/
                            15 hour                   Day-2                   13:15 - 04:16                                                  2000/06/27
\/\/\/\/\/\/\//\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\\/\/\/\/\/\/\\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\//\/\/\/\/\/\/\/\/
2000/06/28
Drk033                                                                                        7200S                                              12H 41M 57S
Drk034                                                                                        7200S                                              19H 37M 23S
Drk035                                                                                        7200S                                              21H 37M 47S
Drk036                                                                                        7200S                                              23H 38M 11S
2000/06/29
Drk037                                                                                        7200S                                              01H 38M 35S
Drk038                                                                                        7200S                                              03H 38M 59S
Drk039                                                                                        7200S                                              05H 39M 23S
PHD040                                                        50000FRs                                                                     10H 41M 54S
 
 
-------------------------------------------------------------------
   File Name                   Pinhole                   PHD                   Dark                   F-F                   Time(start)
-------------------------------------------------------------------
Ana041                                                                                                                 30000FRs             11H 23M 58S
Drk042                                                                                        7200S                                              16H 52M 29S
DEP043                                                                                                                 54000S                  18H 52M 53S
2000/06/30
Drk044                                                                                        7200S                                              09H 53M 17S
Drk045                                                                                        7200S                                              11H 53M 40S
Drk046                                                                                        7200S                                              13H 54M 03S
PHD047                                                        50000FRs                                                                     17H 24M 09S
\/\/\/\/\/\/\\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/
                            70 hour                   Day-3                   18:05 - 16:05                                                  2000/06/30
\/\/\/\/\/\/\//\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\\/\/\/\/\/\/\\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\/\//\/\/\/\/\/\/\/\/
2000/07/03
Drk048                                                                                        7200S                                              17H 20M 36S
Drk049                                                                                        7200S                                              19H 21M 01S
Drk050                                                                                        7200S                                              21H 21M 25S
Drk051                                                                                        7200S                                              23H 21M 49S
2000/07/04
Drk052                                                                                        7200S                                              01H 22M 13S
Drk053                                                                                        7200S                                              03H 22M 37S
Drk054                                                                                        7200S                                              05H 23M 01S
Drk055                                                                                        7200S                                              07H 23M 25S
Drk056                                                                                        7200S                                              09H 23M 49S
Drk057                                                                                        7200S                                              11H 24M 13S
Drk058                                                                                        7200S                                              13H 24M 37S
Drk059                                                                                        7200S                                              15H 25M 01S
DEP060                                                                                                                 54000S                  19H 26M 24S
2000/07/05
Drk061                                                                                        7200S                                              10H 26M 48S
Drk062                                                                                        7200S                                              12H 27M 12S
Ana063                                                                                                                 30000FRs             15H 47M 40S
2000/07/06
PHD064                                                        50000FRs                                                                     09H 39M 36S
-------------------------------------------------------------------------------